Distribution Simulator

The whole molecular weight distribution your polymerization should produce, and the trace your GPC would show of it

Why a shape and not a number

The dispersity predictor tells you what Đ a controlled polymerization should reach. That is the right number to design against, but it is a summary, and two very different reactions can summarize to the same one. A living population at Đ = 1.24 and a tight population at Đ = 1.05 carrying a fifth of its chains dead both read 1.24 on a report and look nothing alike on a trace. One is a slow-exchange problem you fix with a better RAFT agent; the other is a termination problem you fix with lower radical flux. The single number cannot tell you which you have. The shape can.

This page builds the distribution from three pieces, each with a closed form behind it, and then shows you what the instrument would do to it.

Simulate a distribution

The polymerization

C in the equation is 1/Cex. Above ~20 gives Đ near 1.05; below 1 means the agent is mismatched to the monomer.

What went wrong, and what the instrument adds

As a fraction of all chains, not of mass. Dead chains are short, so they weigh far less than they count.

This is how you measure your own band broadening, and it is the better of the two routes because it catches the injector, the detector cell and the tubing as well as the packing. If you have not run one, the trace analyzer estimates the floor from your column series instead, using the guaranteed plate count and the calibration slope.

Hand it to the trace analyzer

This converts the simulated distribution into the two-column chromatogram a GPC would have exported, using log₁₀M = 10.5 − 0.70·V. Paste it into the trace analyzer to see what integrating it would have reported. That closes the round trip: a distribution you know, turned into a trace, integrated back into numbers that no longer quite match.

Three things the shape tells you that Đ does not

Termination hides from Mw. Take a RAFT polymerization sitting at Đ = 1.05 and kill a tenth of its chains. Đ goes to 1.131 and Mₙ falls by 8.1%, but Mw moves only 1.0%, because the dead chains are short and weigh almost nothing. If you are tracking a reaction by Mw, a termination problem is close to invisible; it shows up in Mₙ and in Đ, which is a good reason to report all three rather than the one the instrument prints largest.

Band broadening multiplies Đ. Convolving with a Gaussian in log molecular weight is exact for a log-normal distribution, and it gives a closed form: Đobserved = Đtrue · eσ², where σ is the broadening width in ln M. The factor does not depend on the sample at all, which is why you can measure it once with a standard and apply it to everything. What it does to your conclusions, though, depends entirely on how narrow the sample is:

True ĐObserved ĐTrue Đ − 1Observed Đ − 1Excess inflated by
1.021.0620.0200.062208%
1.051.0930.0500.09386%
1.101.1450.1000.14545%
1.501.5610.5000.56112%
2.002.0821.0001.0828%

That table uses σ = 0.20 in ln M, a middling bank. The dispersity itself is inflated by the same 4.1% in every row, but the quantity you reason about is the excess over 1, and that is inflated by 208% for a genuinely narrow polymer and 8% for a broad one. Band broadening is a narrow-sample problem. It is also why a commercial Đ = 1.02 standard is not really telling you the column is that good, and why chasing Đ below about 1.05 by conventional calibration is chasing your own instrument.

A dead-chain tail and a broad living population are not the same distribution. Set the dispersity to 1.24 two ways, once by dropping Cex to about 4 and once by killing 20% of the chains at Cex = 20. The summary statistic is the same while the curves separate clearly: the first is a single symmetric peak that got wider, the second is a tight peak with a shoulder running down to low molecular weight. On a real trace that shoulder is the thing to look for, and it is the reason to keep the low-molecular-weight end inside your integration limits rather than clipping it off, which the trace analyzer shows the cost of.

What this model does not include

Related tools